SQUID stands for Superconducting Quantum Interference Device. This imaging modality uses a very low-power magnetic field with sensitive detectors that measure the interference of iron within the field. The sensor requires a cryogenic environment, since it must be superconducting to operate. Although SQUID is still considered investigational, linear correlations have been demonstrated between SQUID measurements and liver biopsy LIC levels (1-4).
A measurement is performed by lowering the patient into a known, constant magnetic field, and then detecting the change in magnetic flux versus the change in a water reference medium.
Although SQUID directly measures the magnetic susceptibility of ferritin and hemosiderin, at present it does not have sufficient spatial or temporal resolution to evaluate myocardial iron. In a large clinical trial, LIC data obtained by SQUID were shown to underestimate LIC values obtained from biopsy by a factor of 0.46 (6).
Correlation between SQUID and liver biopsySQUID is a noninvasive test for iron stores that may be frequently repeated, and has been shown to have a linear correlation with LIC assessed by biopsy.
The limited availability and high cost of this test currently restrict its use to research. In addition at present, it appears to underestimate LIC values versus liver biopsy.
The development of liver fibrosis
directly correlates
to liver iron concentrations and serum ferritin levels.